Pulse oximetry












Pulse oximetry
Medical diagnostics

Wrist-oximeter.jpg
A wrist mounted remote sensor pulse oximeter with plethysmograph.

Purpose monitoring a person's oxygen saturation

Pulse oximetry is a noninvasive method for monitoring a person's oxygen saturation (SO2). Though its reading of peripheral oxygen saturation (SpO2) is not always identical to the more desirable reading of arterial oxygen saturation (SaO2) from arterial blood gas analysis, the two are correlated well enough that the safe, convenient, noninvasive, inexpensive pulse oximetry method is valuable for measuring oxygen saturation in clinical use.


In its most common (transmissive) application mode, a sensor device is placed on a thin part of the patient's body, usually a fingertip or earlobe, or in the case of an infant, across a foot. The device passes two wavelengths of light through the body part to a photodetector. It measures the changing absorbance at each of the wavelengths, allowing it to determine the absorbances due to the pulsing arterial blood alone, excluding venous blood, skin, bone, muscle, fat, and (in most cases) nail polish.[1]


Reflectance pulse oximetry is a less common alternative to transmissive pulse oximetery. This method does not require a thin section of the person's body and is therefore well suited to a universal application such as the feet, forehead, and chest, but it also has some limitations. Vasodilation and pooling of venous blood in the head due to compromised venous return to the heart can cause a combination of arterial and venous pulsations in the forehead region and lead to spurious SpO2 results. Such conditions occur while undergoing anesthesia with endotracheal intubation and mechanical ventilation or in patients in the Trendelenburg position.[2]




Contents






  • 1 History


  • 2 Function


  • 3 Indication


    • 3.1 Advantages


    • 3.2 Limitations


    • 3.3 Increasing usage




  • 4 Derived measurements


  • 5 See also


  • 6 Notes


  • 7 References


  • 8 External links





History


In 1935, Karl Matthes (German physician 1905–1962) developed the first 2-wavelength ear O2 saturation meter with red and green filters (later switched to red and infrared filters). His meter was the first device to measure O2 saturation.[3]


The original oximeter was made by Glenn Allan Millikan in the 1940s.[4] In 1949 Wood added a pressure capsule to squeeze blood out of the ear so as to obtain an absolute O2 saturation value when blood was readmitted. The concept is similar to today's conventional pulse oximetry, but was difficult to implement because of unstable photocells and light sources; the method is not now used clinically. In 1964 Shaw assembled the first absolute reading ear oximeter by using eight wavelengths of light.


Pulse oximetry was developed in 1972, by Takuo Aoyagi and Michio Kishi, bioengineers, at Nihon Kohden using the ratio of red to infrared light absorption of pulsating components at the measuring site. Susumu Nakajima, a surgeon, and his associates first tested the device in patients, reporting it in 1975.[5] It was commercialized by Biox in 1980.[6][5][7]


By 1987, the standard of care for the administration of a general anesthetic in the U.S. included pulse oximetry. From the operating room, the use of pulse oximetry rapidly spread throughout the hospital, first to the recovery room, and then into the various intensive care units. Pulse oximetry was of particular value in the neonatal unit where the patients do not thrive with inadequate oxygenation, but too much oxygen and fluctuations in oxygen concentration can lead to vision impairment or blindness from retinopathy of prematurity (ROP). Furthermore, obtaining an arterial blood gas from a neonatal patient is painful to the patient and a major cause of neonatal anemia.[8] Motion artifact can be a significant limitation to pulse oximetry monitoring resulting in frequent false alarms and loss of data. The reason for this is that during motion and low peripheral perfusion, many pulse oximeters cannot distinguish between pulsating arterial blood and moving venous blood, leading to underestimation of oxygen saturation. Early studies of pulse oximetry performance during subject motion made clear the vulnerabilities of conventional pulse oximetry technologies to motion artifact.[9][10]


In 1995, Masimo introduced Signal Extraction Technology (SET) that could measure accurately during patient motion and low perfusion by separating the arterial signal from the venous and other signals. Since then, pulse oximetry manufacturers have developed new algorithms to reduce some false alarms during motion[11] such as extending averaging times or freezing values on the screen, but they do not claim to measure changing conditions during motion and low perfusion. So, there are still important differences in performance of pulse oximeters during challenging conditions.[12] Also in 1995, Masimo introduced perfusion index, quantifying the amplitude of the peripheral plethysmograph waveform. Perfusion index has been shown to help clinicians predict illness severity and early adverse respiratory outcomes in neonates,[13][14][15] predict low superior vena cava flow in very low birth weight infants,[16] provide an early indicator of sympathectomy after epidural anesthesia,[17] and improve detection of critical congenital heart disease in newborns.[18]


Published papers have compared signal extraction technology to other pulse oximetry technologies and have demonstrated consistently favorable results for signal extraction technology.[9][12][19] Signal extraction technology pulse oximetry performance has also been shown to translate into helping clinicians improve patient outcomes. In one study, retinopathy of prematurity (eye damage) was reduced by 58% in very low birth weight neonates at a center using signal extraction technology, while there was no decrease in retinopathy of prematurity at another center with the same clinicians using the same protocol but with non-signal extraction technology.[20] Other studies have shown that signal extraction technology pulse oximetry results in fewer arterial blood gas measurements, faster oxygen weaning time, lower sensor utilization, and lower length of stay.[21] The measure-through motion and low perfusion capabilities it has also allow it to be used in previously unmonitored areas such as the general floor, where false alarms have plagued conventional pulse oximetry. As evidence of this, a landmark study was published in 2010 showing clinicians using signal extraction technology pulse oximetry on the general floor were able to decrease rapid response team activations, ICU transfers, and ICU days.[22]


In 2007, Masimo introduced the first measurement of the pleth variability index (PVI), which multiple clinical studies have shown provides a new method for automatic, noninvasive assessment of a patient's ability to respond to fluid administration.[23][24][25] Appropriate fluid levels are vital to reducing postoperative risks and improving patient outcomes: fluid volumes that are too low (under-hydration) or too high (over-hydration) have been shown to decrease wound healing and increase the risk of infection or cardiac complications.[26] Recently, the National Health Service in the United Kingdom and the French Anesthesia and Critical Care Society listed PVI monitoring as part of their suggested strategies for intra-operative fluid management.[27][28]


In 2011, an expert workgroup recommended newborn screening with pulse oximetry to increase the detection of critical congenital heart disease (CCHD).[29] The CCHD workgroup cited the results of two large, prospective studies of 59,876 subjects that exclusively used signal extraction technology to increase the identification of CCHD with minimal false positives.[30][31] The CCHD workgroup recommended newborn screening be performed with motion tolerant pulse oximetry that has also been validated in low perfusion conditions. In 2011, the US Secretary of Health and Human Services added pulse oximetry to the recommended uniform screening panel.[32] Before the evidence for screening using signal extraction technology, less than 1% of newborns in the United States were screened. Today, The Newborn Foundation has documented near universal screening in the United States and international screening is rapidly expanding.[33] In 2014, a third large study of 122, 738 newborns that also exclusively used signal extraction technology showed similar, positive results as the first two large studies.[34]


High-resolution pulse oximetry (HRPO) has been developed for in-home sleep apnea screening and testing in patients for whom it is impractical to perform polysomnography.[35][36] It stores and records both pulse rate and SpO2 in 1 second intervals and has been shown in one study to help to detect sleep disordered breathing in surgical patients.[37]



Function




Absorption spectra of oxygenated hemoglobin (HbO2) and deoxygenated hemoglobin (Hb) for red and infrared wavelengths


A blood-oxygen monitor displays the percentage of blood that is loaded with oxygen. More specifically, it measures what percentage of hemoglobin, the protein in blood that carries oxygen, is loaded. Acceptable normal ranges for patients without pulmonary pathology are from 95 to 99 percent. For a patient breathing room air at or near sea level, an estimate of arterial pO2 can be made from the blood-oxygen monitor "saturation of peripheral oxygen" (SpO2) reading.


A typical pulse oximeter uses an electronic processor and a pair of small light-emitting diodes (LEDs) facing a photodiode through a translucent part of the patient's body, usually a fingertip or an earlobe. One LED is red, with wavelength of 660 nm, and the other is infrared with a wavelength of 940 nm. Absorption of light at these wavelengths differs significantly between blood loaded with oxygen and blood lacking oxygen. Oxygenated hemoglobin absorbs more infrared light and allows more red light to pass through. Deoxygenated hemoglobin allows more infrared light to pass through and absorbs more red light. The LEDs sequence through their cycle of one on, then the other, then both off about thirty times per second which allows the photodiode to respond to the red and infrared light separately and also adjust for the ambient light baseline.[38]


The amount of light that is transmitted (in other words, that is not absorbed) is measured, and separate normalized signals are produced for each wavelength. These signals fluctuate in time because the amount of arterial blood that is present increases (literally pulses) with each heartbeat. By subtracting the minimum transmitted light from the transmitted light in each wavelength, the effects of other tissues are corrected for, generating a continuous signal for pulsatile arterial blood.[39] The ratio of the red light measurement to the infrared light measurement is then calculated by the processor (which represents the ratio of oxygenated hemoglobin to deoxygenated hemoglobin), and this ratio is then converted to SpO2 by the processor via a lookup table[39] based on the Beer–Lambert law.[38] The signal separation also serves other purposes: a plethysmograph waveform ("pleth wave") representing the pulsatile signal is usually displayed for a visual indication of the pulses as well as signal quality,[40] and a numeric ratio between the pulsatile and baseline absorbance ("perfusion index") can be used to evaluate perfusion.[24]



Indication




A pulse oximeter probe applied to a person's finger


A pulse oximeter is a medical device that indirectly monitors the oxygen saturation of a patient's blood (as opposed to measuring oxygen saturation directly through a blood sample) and changes in blood volume in the skin, producing a photoplethysmogram that may be further processed into other measurements.[40] The pulse oximeter may be incorporated into a multiparameter patient monitor. Most monitors also display the pulse rate. Portable, battery-operated pulse oximeters are also available for transport or home blood-oxygen monitoring.



Advantages


Pulse oximetry is particularly convenient for noninvasive continuous measurement of blood oxygen saturation. In contrast, blood gas levels must otherwise be determined in a laboratory on a drawn blood sample. Pulse oximetry is useful in any setting where a patient's oxygenation is unstable, including intensive care, operating, recovery, emergency and hospital ward settings, pilots in unpressurized aircraft, for assessment of any patient's oxygenation, and determining the effectiveness of or need for supplemental oxygen. Although a pulse oximeter is used to monitor oxygenation, it cannot determine the metabolism of oxygen, or the amount of oxygen being used by a patient. For this purpose, it is necessary to also measure carbon dioxide (CO2) levels. It is possible that it can also be used to detect abnormalities in ventilation. However, the use of a pulse oximeter to detect hypoventilation is impaired with the use of supplemental oxygen, as it is only when patients breathe room air that abnormalities in respiratory function can be detected reliably with its use. Therefore, the routine administration of supplemental oxygen may be unwarranted if the patient is able to maintain adequate oxygenation in room air, since it can result in hypoventilation going undetected.[41]


Because of their simplicity of use and the ability to provide continuous and immediate oxygen saturation values, pulse oximeters are of critical importance in emergency medicine and are also very useful for patients with respiratory or cardiac problems, especially COPD, or for diagnosis of some sleep disorders such as apnea and hypopnea.[42] Portable battery-operated pulse oximeters are useful for pilots operating in a non-pressurized aircraft above 10,000 feet (3,000 m) or 12,500 feet (3,800 m) in the U.S.[43] where supplemental oxygen is required. Portable pulse oximeters are also useful for mountain climbers and athletes whose oxygen levels may decrease at high altitudes or with exercise. Some portable pulse oximeters employ software that charts a patient's blood oxygen and pulse, serving as a reminder to check blood oxygen levels.



Limitations


Pulse oximetry solely measures hemoglobin saturation, not ventilation and is not a complete measure of respiratory sufficiency. It is not a substitute for blood gases checked in a laboratory, because it gives no indication of base deficit, carbon dioxide levels, blood pH, or bicarbonate (HCO3) concentration. The metabolism of oxygen can be readily measured by monitoring expired CO2, but saturation figures give no information about blood oxygen content. Most of the oxygen in the blood is carried by hemoglobin; in severe anemia, the blood contains less hemoglobin, which despite being saturated cannot carry as much oxygen.


Erroneously low readings may be caused by hypoperfusion of the extremity being used for monitoring (often due to a limb being cold, or from vasoconstriction secondary to the use of vasopressor agents); incorrect sensor application; highly calloused skin; or movement (such as shivering), especially during hypoperfusion. To ensure accuracy, the sensor should return a steady pulse and/or pulse waveform. Pulse oximetry technologies differ in their abilities to provide accurate data during conditions of motion and low perfusion.[12][9]


Pulse oximetry also is not a complete measure of circulatory sufficiency. If there is insufficient bloodflow or insufficient hemoglobin in the blood (anemia), tissues can suffer hypoxia despite high oxygen saturation in the blood that does arrive.


Since pulse oximetry measures only the percentage of bound hemoglobin, a falsely high or falsely low reading will occur when hemoglobin binds to something other than oxygen:



  • Hemoglobin has a higher affinity to carbon monoxide than it does to oxygen, and a high reading may occur despite the patient's actually being hypoxemic. In cases of carbon monoxide poisoning, this inaccuracy may delay the recognition of hypoxia (low cellular oxygen level).


  • Cyanide poisoning gives a high reading because it reduces oxygen extraction from arterial blood. In this case, the reading is not false, as arterial blood oxygen is indeed high in early cyanide poisoning.


  • Methemoglobinemia characteristically causes pulse oximetry readings in the mid-80s.

  • COPD [especially chronic bronchitis] may cause false readings.[44]


A noninvasive method that allows continuous measurement of the dyshemoglobins is the pulse CO-oximeter, which was built in 2005 by Masimo.[45] By using additional wavelengths,[46] it provides clinicians a way to measure the dyshemoglobins, carboxyhemoglobin, and methemoglobin along with total hemoglobin.[47]



Increasing usage


According to a report by iData Research the U.S. pulse oximetry monitoring market for equipment and sensors was over 700 million USD in 2011.[48]


In 2008, more than half of the major internationally exporting medical equipment manufacturers in China were producers of pulse oximeters.[49]


The Apple Watch uses this technology for its heart rate monitor.[50][51] The accuracy of the heart rate monitor has been debated, as different tests have shown it to be 91% as accurate as chest strap monitors.[52][53]



Derived measurements



Due to changes in blood volumes in the skin, a plethysmographic variation can be seen in the light signal received (transmittance) by the sensor on an oximeter. The variation can be described as a periodic function, which in turn can be split into a DC component (the peak value)[a] and an AC component (peak minus valley).[54] The ratio of the AC component to the DC component, expressed as a percentage, is known as the (peripheral) perfusion index (Pi) fpr a pulse, and typically has a range of 0.02% to 20%.[55] An earlier measurement called the pulse oximetry plethysmographic (POP) only measures the "AC" component, and is derived manually from monitor pixels.[56][24]


Pleth variability index (PVI) is a measure of the variability of the perfusion index, which occurs during breathing cycles. Mathematically it is calculated as (Pimax - Pimin)/Pimax × 100%, where the maximum and minimum Pi values are from one or many breathing cycles.[54] It has been shown to be a useful, noninvasive indicator of continuous fluid responsiveness for patients undergoing fluid management.[24]Pulse oximetry plethysmographic waveform amplitude (ΔPOP) is an analogous earlier technique for use on the manually-derived POP, calculated as (POPmax - POPmin)/(POPmax + POPmin)*2.[56]



See also




  • Arterial blood gas

  • Capnography

  • Integrated Pulmonary Index

  • Respiratory monitoring

  • Medical equipment

  • Mechanical ventilation

  • Oxygen sensor

  • Oxygen saturation


  • Photoplethysmogram, measuring of carbon dioxide (CO2) in the respiratory gases

  • Sleep apnea




Notes





  1. ^ This definition used by Masimo varies from the mean value used in signal processing; it is meant to measure the pulsatile arterial blood absorbance over the baseline absorbance.




References





  1. ^ Brand TM, Brand ME, Jay GD (February 2002). "Enamel nail polish does not interfere with pulse oximetry among normoxic volunteers". Journal of Clinical Monitoring and Computing. 17 (2): 93–6. doi:10.1023/A:1016385222568. PMID 12212998..mw-parser-output cite.citation{font-style:inherit}.mw-parser-output .citation q{quotes:"""""""'""'"}.mw-parser-output .citation .cs1-lock-free a{background:url("//upload.wikimedia.org/wikipedia/commons/thumb/6/65/Lock-green.svg/9px-Lock-green.svg.png")no-repeat;background-position:right .1em center}.mw-parser-output .citation .cs1-lock-limited a,.mw-parser-output .citation .cs1-lock-registration a{background:url("//upload.wikimedia.org/wikipedia/commons/thumb/d/d6/Lock-gray-alt-2.svg/9px-Lock-gray-alt-2.svg.png")no-repeat;background-position:right .1em center}.mw-parser-output .citation .cs1-lock-subscription a{background:url("//upload.wikimedia.org/wikipedia/commons/thumb/a/aa/Lock-red-alt-2.svg/9px-Lock-red-alt-2.svg.png")no-repeat;background-position:right .1em center}.mw-parser-output .cs1-subscription,.mw-parser-output .cs1-registration{color:#555}.mw-parser-output .cs1-subscription span,.mw-parser-output .cs1-registration span{border-bottom:1px dotted;cursor:help}.mw-parser-output .cs1-ws-icon a{background:url("//upload.wikimedia.org/wikipedia/commons/thumb/4/4c/Wikisource-logo.svg/12px-Wikisource-logo.svg.png")no-repeat;background-position:right .1em center}.mw-parser-output code.cs1-code{color:inherit;background:inherit;border:inherit;padding:inherit}.mw-parser-output .cs1-hidden-error{display:none;font-size:100%}.mw-parser-output .cs1-visible-error{font-size:100%}.mw-parser-output .cs1-maint{display:none;color:#33aa33;margin-left:0.3em}.mw-parser-output .cs1-subscription,.mw-parser-output .cs1-registration,.mw-parser-output .cs1-format{font-size:95%}.mw-parser-output .cs1-kern-left,.mw-parser-output .cs1-kern-wl-left{padding-left:0.2em}.mw-parser-output .cs1-kern-right,.mw-parser-output .cs1-kern-wl-right{padding-right:0.2em}


  2. ^ Jørgensen JS, Schmid ER, König V, Faisst K, Huch A, Huch R (July 1995). "Limitations of forehead pulse oximetry". Journal of Clinical Monitoring. 11 (4): 253–6. doi:10.1007/bf01617520. PMID 7561999.


  3. ^ Matthes K (1935). "Untersuchungen über die Sauerstoffsättigung des menschlichen Arterienblutes" [Studies on the Oxygen Saturation of Arterial Human Blood]. Naunyn-Schmiedeberg's Archives of Pharmacology (in German). 179 (6): 698–711. doi:10.1007/BF01862691.


  4. ^ Millikan GA (1942). "The oximeter: an instrument for measuring continuously oxygen saturation of arterial blood in man". Review of Scientific Instruments. 13 (10): 434–444. Bibcode:1942RScI...13..434M. doi:10.1063/1.1769941.


  5. ^ ab Severinghaus JW, Honda Y (April 1987). "History of blood gas analysis. VII. Pulse oximetry". Journal of Clinical Monitoring. 3 (2): 135–8. doi:10.1007/bf00858362. PMID 3295125.


  6. ^ "510(k): Premarket Notification". United States Food and Drug Administration. Retrieved 2017-02-23.


  7. ^ "Fact vs. Fiction". Masimo Corporation. Archived from the original on 13 April 2009.


  8. ^ Lin JC, Strauss RG, Kulhavy JC, Johnson KJ, Zimmerman MB, Cress GA, Connolly NW, Widness JA (August 2000). "Phlebotomy overdraw in the neonatal intensive care nursery". Pediatrics. 106 (2): E19. doi:10.1542/peds.106.2.e19. PMID 10920175.


  9. ^ abc Barker SJ (October 2002). ""Motion-resistant" pulse oximetry: a comparison of new and old models". Anesthesia and Analgesia. 95 (4): 967–72. doi:10.1213/00000539-200210000-00033. PMID 12351278.


  10. ^ Barker SJ, Shah NK (October 1996). "Effects of motion on the performance of pulse oximeters in volunteers". Anesthesiology. 85 (4): 774–81. doi:10.1097/00000542-199701000-00014. PMID 8873547.


  11. ^ Jopling MW, Mannheimer PD, Bebout DE (January 2002). "Issues in the laboratory evaluation of pulse oximeter performance". Anesthesia and Analgesia. 94 (1 Suppl): S62–8. PMID 11900041.


  12. ^ abc Shah N, Ragaswamy HB, Govindugari K, Estanol L (August 2012). "Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers". Journal of Clinical Anesthesia. 24 (5): 385–91. doi:10.1016/j.jclinane.2011.10.012. PMID 22626683.


  13. ^ De Felice C, Leoni L, Tommasini E, Tonni G, Toti P, Del Vecchio A, Ladisa G, Latini G (March 2008). "Maternal pulse oximetry perfusion index as a predictor of early adverse respiratory neonatal outcome after elective cesarean delivery". Pediatric Critical Care Medicine. 9 (2): 203–8. doi:10.1097/pcc.0b013e3181670021. PMID 18477934.


  14. ^ De Felice C, Latini G, Vacca P, Kopotic RJ (October 2002). "The pulse oximeter perfusion index as a predictor for high illness severity in neonates". European Journal of Pediatrics. 161 (10): 561–2. doi:10.1007/s00431-002-1042-5. PMID 12297906.


  15. ^ De Felice C, Goldstein MR, Parrini S, Verrotti A, Criscuolo M, Latini G (March 2006). "Early dynamic changes in pulse oximetry signals in preterm newborns with histologic chorioamnionitis". Pediatric Critical Care Medicine. 7 (2): 138–42. doi:10.1097/01.PCC.0000201002.50708.62. PMID 16474255.


  16. ^ Takahashi S, Kakiuchi S, Nanba Y, Tsukamoto K, Nakamura T, Ito Y (April 2010). "The perfusion index derived from a pulse oximeter for predicting low superior vena cava flow in very low birth weight infants". Journal of Perinatology. 30 (4): 265–9. doi:10.1038/jp.2009.159. PMC 2834357. PMID 19907430.


  17. ^ Ginosar Y, Weiniger CF, Meroz Y, Kurz V, Bdolah-Abram T, Babchenko A, Nitzan M, Davidson EM (September 2009). "Pulse oximeter perfusion index as an early indicator of sympathectomy after epidural anesthesia". Acta Anaesthesiologica Scandinavica. 53 (8): 1018–26. doi:10.1111/j.1399-6576.2009.01968.x. PMID 19397502.


  18. ^ Granelli A, Ostman-Smith I (October 2007). "Noninvasive peripheral perfusion index as a possible tool for screening for critical left heart obstruction". Acta Paediatrica. 96 (10): 1455–9. doi:10.1111/j.1651-2227.2007.00439.x. PMID 17727691.


  19. ^ Hay WW, Rodden DJ, Collins SM, Melara DL, Hale KA, Fashaw LM (2002). "Reliability of conventional and new pulse oximetry in neonatal patients". Journal of Perinatology. 22 (5): 360–6. doi:10.1038/sj.jp.7210740. PMID 12082469.


  20. ^ Castillo A, Deulofeut R, Critz A, Sola A (February 2011). "Prevention of retinopathy of prematurity in preterm infants through changes in clinical practice and SpO₂technology". Acta Paediatrica. 100 (2): 188–92. doi:10.1111/j.1651-2227.2010.02001.x. PMC 3040295. PMID 20825604.


  21. ^ Durbin CG, Rostow SK (August 2002). "More reliable oximetry reduces the frequency of arterial blood gas analyses and hastens oxygen weaning after cardiac surgery: a prospective, randomized trial of the clinical impact of a new technology". Critical Care Medicine. 30 (8): 1735–40. doi:10.1097/00003246-200208000-00010. PMID 12163785.


  22. ^ Taenzer AH, Pyke JB, McGrath SP, Blike GT (February 2010). "Impact of pulse oximetry surveillance on rescue events and intensive care unit transfers: a before-and-after concurrence study". Anesthesiology. 112 (2): 282–7. doi:10.1097/aln.0b013e3181ca7a9b. PMID 20098128.


  23. ^ Zimmermann M, Feibicke T, Keyl C, Prasser C, Moritz S, Graf BM, Wiesenack C (June 2010). "Accuracy of stroke volume variation compared with pleth variability index to predict fluid responsiveness in mechanically ventilated patients undergoing major surgery". European Journal of Anaesthesiology. 27 (6): 555–61. doi:10.1097/EJA.0b013e328335fbd1. PMID 20035228.


  24. ^ abcd Cannesson M, Desebbe O, Rosamel P, Delannoy B, Robin J, Bastien O, Lehot JJ (August 2008). "Pleth variability index to monitor the respiratory variations in the pulse oximeter plethysmographic waveform amplitude and predict fluid responsiveness in the operating theatre". British Journal of Anaesthesia. 101 (2): 200–6. doi:10.1093/bja/aen133. PMID 18522935.


  25. ^ Forget P, Lois F, de Kock M (October 2010). "Goal-directed fluid management based on the pulse oximeter-derived pleth variability index reduces lactate levels and improves fluid management". Anesthesia and Analgesia. 111 (4): 910–4. doi:10.1213/ANE.0b013e3181eb624f. PMID 20705785.


  26. ^ Ishii M, Ohno K (March 1977). "Comparisons of body fluid volumes, plasma renin activity, hemodynamics and pressor responsiveness between juvenile and aged patients with essential hypertension". Japanese Circulation Journal. 41 (3): 237–46. doi:10.1253/jcj.41.237. PMID 870721.


  27. ^ "NHS Technology Adoption Centre". Ntac.nhs.uk. Retrieved 2015-04-02.


  28. ^ Vallet B, Blanloeil Y, Cholley B, Orliaguet G, Pierre S, Tavernier B (October 2013). "Guidelines for perioperative haemodynamic optimization". Annales Francaises d'Anesthesie et de Reanimation. 32 (10): e151–8. doi:10.1016/j.annfar.2013.09.010. PMID 24126197.


  29. ^ Kemper AR, Mahle WT, Martin GR, Cooley WC, Kumar P, Morrow WR, Kelm K, Pearson GD, Glidewell J, Grosse SD, Howell RR (November 2011). "Strategies for implementing screening for critical congenital heart disease". Pediatrics. 128 (5): e1259–67. doi:10.1542/peds.2011-1317. PMID 21987707.


  30. ^ de-Wahl Granelli A, Wennergren M, Sandberg K, Mellander M, Bejlum C, Inganäs L, Eriksson M, Segerdahl N, Agren A, Ekman-Joelsson BM, Sunnegårdh J, Verdicchio M, Ostman-Smith I (January 2009). "Impact of pulse oximetry screening on the detection of duct dependent congenital heart disease: a Swedish prospective screening study in 39,821 newborns". BMJ. 338: a3037. doi:10.1136/bmj.a3037. PMC 2627280. PMID 19131383.


  31. ^ Ewer AK, Middleton LJ, Furmston AT, Bhoyar A, Daniels JP, Thangaratinam S, Deeks JJ, Khan KS (August 2011). "Pulse oximetry screening for congenital heart defects in newborn infants (PulseOx): a test accuracy study". Lancet. 378 (9793): 785–94. doi:10.1016/S0140-6736(11)60753-8. PMC 3860684. PMID 21820732.


  32. ^ Mahle WT, Martin GR, Beekman RH, Morrow WR (January 2012). "Endorsement of Health and Human Services recommendation for pulse oximetry screening for critical congenital heart disease". Pediatrics. 129 (1): 190–2. doi:10.1542/peds.2011-3211. PMID 22201143.


  33. ^ "Newborn CCHD Screening Progress Map". Cchdscreeningmap.org. 7 July 2014. Retrieved 2015-04-02.


  34. ^ Zhao QM, Ma XJ, Ge XL, Liu F, Yan WL, Wu L, Ye M, Liang XC, Zhang J, Gao Y, Jia B, Huang GY (August 2014). "Pulse oximetry with clinical assessment to screen for congenital heart disease in neonates in China: a prospective study". Lancet. 384 (9945): 747–54. doi:10.1016/S0140-6736(14)60198-7. PMID 24768155.


  35. ^ Valenza T (April 2008). "Keeping a Pulse on Oximetry". Archived from the original on February 10, 2012.


  36. ^ "PULSOX -300i" (PDF). Maxtec Inc. Archived from the original (PDF) on January 7, 2009.


  37. ^ Chung F, Liao P, Elsaid H, Islam S, Shapiro CM, Sun Y (May 2012). "Oxygen desaturation index from nocturnal oximetry: a sensitive and specific tool to detect sleep-disordered breathing in surgical patients". Anesthesia and Analgesia. 114 (5): 993–1000. doi:10.1213/ane.0b013e318248f4f5. PMID 22366847.


  38. ^ ab "Principles of pulse oximetry". Anaesthesia UK. 11 Sep 2004. Archived from the original on 2015-02-24. Retrieved 2015-02-24.


  39. ^ ab "Pulse Oximetry". Oximetry.org. 2002-09-10. Archived from the original on 2015-03-18. Retrieved 2015-04-02.


  40. ^ ab "SpO2 monitoring in the ICU" (PDF). Liverpool Hospital. Retrieved 24 March 2019.


  41. ^ Fu ES, Downs JB, Schweiger JW, Miguel RV, Smith RA (November 2004). "Supplemental oxygen impairs detection of hypoventilation by pulse oximetry". Chest. 126 (5): 1552–8. doi:10.1378/chest.126.5.1552. PMID 15539726.


  42. ^ Schlosshan D, Elliott MW (April 2004). "Sleep . 3: Clinical presentation and diagnosis of the obstructive sleep apnoea hypopnoea syndrome". Thorax. 59 (4): 347–52. doi:10.1136/thx.2003.007179. PMC 1763828. PMID 15047962.


  43. ^ "FAR Part 91 Sec. 91.211 effective as of 09/30/1963". Airweb.faa.gov. Retrieved 2015-04-02.


  44. ^ Amalakanti S, Pentakota MR (April 2016). "Pulse Oximetry Overestimates Oxygen Saturation in COPD". Respiratory Care. 61 (4): 423–7. doi:10.4187/respcare.04435. PMID 26715772.


  45. ^ ‹See Tfd›UK 2320566 


  46. ^ Maisel, William; Roger J. Lewis (2010). "Noninvasive Measurement of Carboxyhemoglobin: How Accurate is Accurate Enough?". Annals of Emergency Medicine. 56 (4): 389–91. doi:10.1016/j.annemergmed.2010.05.025. PMID 20646785.


  47. ^ "Total Hemoglobin (SpHb)". Masimo. Retrieved 24 March 2019.


  48. ^ U.S. Market for Patient Monitoring Equipment. iData Research. May 2012


  49. ^ "Key Portable Medical Device Vendors Worldwide". China Portable Medical Devices Report. December 2008.


  50. ^ Lovejoy B (24 April 2015). "Apple Watch teardown reveals pulse oximeter, suggesting future measurement of blood oxygen". 9 to 5 Mac.


  51. ^ "Your heart rate. What it means, and where on Apple Watch you'll find it". Apple.


  52. ^ Ross E (15 October 2016). "You Can Monitor Heart Rhythm With A Smartphone, But Should You?". National Public Radio.


  53. ^ Peloquin A. "Chest Strap Vs Wristband Heart Rate Monitors". Breaking Muscle.


  54. ^ ab U.S. Patent 8,414,499


  55. ^ Lima, A; Bakker, J (October 2005). "Noninvasive monitoring of peripheral perfusion". Intensive care medicine. 31 (10): 1316–26. doi:10.1007/s00134-005-2790-2. PMID 16170543.


  56. ^ ab Cannesson, M; Attof, Y; Rosamel, P; Desebbe, O; Joseph, P; Metton, O; Bastien, O; Lehot, JJ (June 2007). "Respiratory variations in pulse oximetry plethysmographic waveform amplitude to predict fluid responsiveness in the operating room". Anesthesiology. 106 (6): 1105–11. doi:10.1097/01.anes.0000267593.72744.20. PMID 17525584.




External links






  • [1]







Popular posts from this blog

Guess what letter conforming each word

Port of Spain

Run scheduled task as local user group (not BUILTIN)